Software in Medical Devices

As a software developer in MedTech, you face a stark contrast between conventional software engineering and regulated medical device development. While the tech industry celebrates rapid iteration and “move fast and break things,” medical device development prioritizes patient safety, regulatory compliance, and exhaustive documentation. These aren’t optional as they are required before your software can reach a single patient. The challenge extends beyond writing functional code to simultaneously proving to regulators that it’s safe, effective, and secure.

Software verification and validation (V&V) in regulated environments goes beyond executing software testing. As described in ISO 13485, the tools and infrastructure used to develop the software itself and those tools and processes used to get the software into production must also be validated.

Validation must follow documented protocols with predefined acceptance criteria established before testing even begins. Requirements documentation forms the foundation to determine what testing is needed. User Requirements Documentation must specify intended use, operating environment, performance requirements, and safety related functions. Software Requirements Specification must document all inputs, outputs, functions, interfaces, error handling, and safety related features. The design file must include testing protocols with traceability between requirements and tests, clear acceptance criteria, and results with appropriate evidence to demonstrate safety, security, and effectiveness.

The device manufacturer retains ultimate responsibility for validation even when using third party contractors or testing laboratories.